Universal Laser Systems VersaLASER VLS2.30 and VLS3.50 Installation and Setup Guide
Electrical Power Source (User Supplied)
For your systems electrical requirements, please refer to the “INPUT POWER” label near the power inlet.1.
CAUTION:2. Never remove the ground lead to the electrical cord and plug the VLS
system into a non-grounded outlet.
A laser system that is not properly grounded is hazardous and has
the potential to cause severe or fatal electrical shock. Without proper grounding, the laser system may exhibit
sporadic or unpredictable behavior. Always plug the system into a properly grounded (earthed) outlet.
Noisy or unstable electricity and voltage spikes may cause interference and possible damage to the3.
electronics of the laser system. If electrical power  uctuations, brown outs or constant power outages are a
problem in your area, please contact your local Electrician to supply a power isolation and regulation module.
Electrically noisy equipment, such as equipment with large motors, can also cause interference if plugged into
the same outlet. It may be necessary to attach the laser system to a dedicated electrical line to resolve the
problem.
The laser system is designed as a Class I, Group A, pluggable device. It is also designed for connection to IT4.
power systems which provide the most  exibility to the user.
Exhaust System (User Supplied)
We recommend you consult with a licensed contractor to meet local safety and building code requirements.1.
The exhaust system must be capable of supplying a minimum of:2.
VLS2.30 - 150 CFM (cubic feet per minute) of air ow while under a load of 6 inches of static pressure
(254m3/hr at 1.5kPa)
VLS3.50 - 250 CFM (cubic feet per minute) of air ow while under a load of 6 inches of static pressure
(425m3/hr at 1.5kPa)
Do not install forward incline, backward incline, in-line or ventilator fans because these types of air handlers3.
are inadequate and inappropriate for this type of installation. A high-pressure blower must be used to meet
minimum air ow requirements.
For personal safety and noise control, we recommend that the exhaust blower be mounted outside the building.4.
Rigid tubing should be used for the majority of the connection between the blower and the laser system. The5.
tubing should be smooth-walled and have as few 90-degree bends as possible.
Install an exhaust gate to adjust air ow and to close o the exhaust when the laser is not in use. Place this gate 6.
near the laser system within 5 to 10 feet (1.50 to 3.00 meters).
Use a short piece of industrial grade, wire-reinforced rubber tubing to connect the laser systems exhaust port 7.
to the exhaust gate and secure with a hose clamp. This will provide some mobility for your laser system and
will isolate exhaust blower vibrations from your laser system.
Wire the exhaust blower electrically to a wall switch in the same room for easy ON/OFF control.8.